Spray-can penetrant testing agents are characterized by “spray, use, and discard,” as they are sealed under positive pressure and are neither recovered nor reused. Under normal conditions, the penetrant in the spray can remains uncontaminated. Standards such as AMS-2644-2020, ISO 3452-2:2013, GB/T 18851.2-2008, and T/CAMAC 0005-2020 “Nondestructive Testing in Civil Aviation—Penetrant Testing” do not require in-process control inspections of spray-can penetrant testing agents during use.
2.3 Verification of the Sensitivity of the Penetrant Testing System and the Correctness of the Operating Procedures
Verification of the sensitivity of penetrant testing systems and the correctness of their operating procedures is fundamentally different from verification of penetrant sensitivity.
1. Verification of the sensitivity of the penetrant testing system and the correctness of the operating procedures shall be performed using a penetrant whose sensitivity is to be verified, including the developer.
2. Verification of the sensitivity of a penetrant testing system does not require comparison with a reference penetrant; the sensitivity level can be directly determined by the number of crack indications revealed by the penetrant under test.
3. The procedures for verifying the sensitivity of the penetrant testing system and the correctness of the testing procedure are identical to those used by personnel performing penetrant testing. Standards NB/T 20003.4-2010, NB/T 20328.4-2015, and NB/T 47013.5-2015 specify the use of the three-point Type B test block outlined in JB/T 6064-2015 to verify both the sensitivity of the penetrant testing system and the correctness of the testing procedure.
To broaden the measurable sensitivity range, a five-point Type B test block as shown in Figure 5 may also be used (from top to bottom, the “+” symbols denote crack regions 1, 2, 3, 4, and 5, respectively). The three crack regions on the three-point Type B test block roughly correspond to crack regions 2, 3, and 4 on the five-point Type B test block.
Figure 4: Physical specimen of Type 2 reference test block and its structural schematic
According to these two standards, the test block is used for process control inspection of penetrant testing systems; however, process control inspection is not applicable to spray-can penetrant testing systems.
Figure 5 Schematic Diagram of the Five-Point Type B Test Block Configuration
Although the Type 2 reference test block specified in ISO 3452-3:2013 and GB/T 18851.3-2008 also comprises five crack indications, their maximum dimensions range from 3.0 to 5.5 mm, which is narrower than the 1.0–6.0 mm range of the five-point Type B test block. The smallest crack indication on the Type 2 reference test block roughly corresponds to the third crack indication on the five-point Type B test block (i.e., the second crack indication on the three-point Type B test block). Compared with the five-point Type B test block, the Type 2 reference test block offers a narrower sensitivity range and a lower maximum detectable sensitivity. Even when all five crack indications on the Type 2 reference test block are clearly visible, the achievable sensitivity level is only that of Class B as specified in NB/T 47013.5-2015; it thus fails to meet the requirements for penetrant testing system sensitivity set forth in NB/T 20003.4-2010 and NB/T 20328.4-2015. If verification of a higher sensitivity class is required, the PSM-5 test block shown in Figure 6 may be used. This test block is widely employed in process-control inspections for penetrant testing in the civil aviation sector, with its smallest crack indication having a major axis length of 0.38 to 0.79 mm.
Figure 6: Appearance of the PSM-5 Test Block
Whether it is Type B test blocks, Type 2 reference test blocks, or PSM-5 test blocks, they all have the following drawbacks:
1. Cracks on the test specimen are naturally formed under compressive loading applied to the back surface of the specimen. During service, stress inputs such as impact, drop, bending loads, or additional pressure on the back surface can cause these cracks to propagate, increase in size, and even give rise to new cracks.
2. Cracks on the test block have extremely sharp tips, which continuously accumulate penetrant during use, causing the indications for these fine cracks to become increasingly faint or even disappear altogether as the number of inspections increases. This issue can be addressed by ultrasonic cleaning of the test block, but the results are not always satisfactory; ultimately, the test block must still be replaced.
3. The surface roughness of the crack zone on the test block generally does not exceed 2.5 μm, which is superior to almost all surfaces to be inspected. This results in penetrant being very easily removed from the smooth surface of the test block; if the same overpenetrant removal procedure used for the workpiece under inspection is applied to the test block, there is a high likelihood of overcleaning the test block.
4. Cracks on the test blocks form naturally under pressure; therefore, the width and depth of the cracks cannot be precisely controlled, and no two test blocks are exactly alike. This may result in penetration testing agents with the same system sensitivity yielding different sensitivity ratings when applied to different test blocks.
The sensitivity of penetrant testing agents to a certain extent limits the upper bound of sensitivity that penetrant testing can achieve. To ensure that the sensitivity of penetrant testing meets the required standards, the minimum sensitivity grade for penetrant testing agents should be specified during implementation. Furthermore, the sensitivity grade of penetrant testing agents shall be self-declared by the supplier, accompanied by a sensitivity certification report issued by an accredited third-party laboratory as supporting documentation. It is recommended that the relevant authorities, following the model of standard QPL-AMS-2644-2020, establish a national catalog of certified and qualified penetrant testing materials to facilitate the procurement of penetrant testing agents that meet the required sensitivity grades. Verification of the system sensitivity of penetrant testing agents and the correctness of the application procedures is suitable for promotion within the penetrant testing industry; however, using Type B test blocks, Type 2 reference test blocks, or PSM-5 test blocks for such verification yields only a single measure of sensitivity, which does not correspond directly to actual defect detection capability and should therefore be regarded merely as a reference.